Using the Stellar device for the first time
When using the Stellar device for the first time, ensure that all components are in working
condition and perform a functional test (see “Performing a functional test” on page 13). Alarms
should also be checked by the physician when setting up a new patient (see Clinical Guide).
Working with other optional accessories
Attaching the pulse oximeter
Use of a pulse oximeter may be recommended by your clinician.
Contraindication
The pulse oximeter does not meet defibrillation-proof requirement per IEC 60601-1: 1990,
clause 17.h.
1
Connect the plug of the finger pulse sensor to the plug of the pulse oximeter.
2
Connect the plug of the pulse oximeter at the rear of the device.
To view the oximetry values, from the
Monitoring
menu, select
Monitoring
.
Adding supplemental oxygen
Oxygen may be prescribed by your clinician.
Note:
Up to 30 L/min can be added.
WARNING
•
Oxygen flow must be turned off when the device is not operating, so that unused
oxygen does not accumulate within the device and create a risk of fire.
•
Use only certified, clean oxygen sources.
•
ResMed strongly recommends adding oxygen into Stellar’s oxygen inlet at the
rear of the device. Entraining oxygen elsewhere, ie into the breathing system via
a side port or at the mask, has potential to impair triggering and accuracy of
therapy/monitoring and alarms (eg, High Leak alarm, Non-vented mask alarm). If
used in this way, therapy and alarm operation must be verified each time oxygen
flow is adjusted.
•
Oxygen supports combustion. Oxygen must not be used while smoking or in the
presence of an open flame. Only use oxygen supply in well-ventilated rooms.
•
The breathing system and the oxygen source must be kept at a minimum
distance of 2 m away from any sources of ignition (eg, electrical devices).
•
Oxygen must not be used while the device is being operated within the mobility bag.
1
2