NOT013281-8
6.5 Alarm testing procedures
CAUTION
The visual and audible signals of all alarms must be tested regularly to ensure they are
working properly, particularly:
•
Prior to connection to a patient, before continuous ventilation
•
After any configuration or alarm setting change.
Note:
Some alarms may have a reaction time of several seconds before they are activated or
deactivated.
Testing the power supply alarms
1. Connect a patient circuit (double or single) fitted with a Maquet test lung to the ventilator and
perform the manual test.
2. Set the mode to ACV, Adult.
3. Mute any alarms which may be triggered (by pressing the Alarm Silence button twice).
4. Perform the following test for each power supply type (mains, external DC and external
battery):
Mini Vte (mL)
10
300
10 / NO
NO / 10
500
1000
10
1000
2300
100
Maxi Vti (mL)
(Pressure-
controlled modes)
750
550
1000
50
200
10
200
500
750 / NO
1000
50
1000
2400 / NO
100
Maxi Vti (mL)
(Volume-controlled
modes)
750
550
1000
50
50
200
10
200
200
750 / NO
1000
50
1000
2400 / NO
100
Mini Vti (mL)
10
300
10 / NO
10 / NO
500
1000
10
1000
2300
100
Δ Pi (cm H
2
O)
3
1
20
1
High Leak (%)
NO
20
80 / NO
1
Default value
Settings range
Value:
Minimum
Maximum
Increment
Patient type:
Child
Adult
Child
Adult
Child
Adult
Action
Visual alert
Audible signal
Connect the power supply
Off
Off
Disconnect the power supply
Flashing
Sounding
Reconnect the power supply
Steady
Off
Press the Alarm Silence button
Off
Off
Summary of Contents for Elisee 350
Page 1: ...Elis e 350 Clinical Manual English Making quality of care easy Respiratory Care solutions...
Page 2: ...Making quality of care easy Respiratory Care solutions...
Page 4: ......
Page 6: ...iv...
Page 10: ...4 102 NOT013281 8...
Page 20: ...14 102 NOT013281 8...
Page 50: ...44 102 NOT013281 8...
Page 86: ...80 102 NOT013281 8...
Page 109: ...Making quality of care easy Respiratory Care solutions...