5.2 Starting up
When the ventilator is running on battery power
1. Press the
button on the right-hand side of the device. The device performs its self-test,
which includes:
•
Checking available power supplies
•
Setting the flow sensor offsets
•
Displaying the language of the software using the ISO 639-1 codes:
(EN: English, DE: German, FR: French, IT: Italian, ES: Spanish, SV: Swedish, NO:
Norwegian).
•
When the self-test is complete, the device displays the main screen:
When the ventilator is running on mains power
As soon as the device is connected to the mains supply, it conducts its self-test. You can then
list the available power supplies, set the flow sensor offsets, and test the occlusion pressure
calculation function (P0.1) (see “Respiratory mechanics parameters” on page 58). Next, the
screen showing the internal and external battery charge indicator will be displayed.
1. Press the
button on the right-hand of the device. The main screen is displayed.
5.3 Checks before use
Before the ventilator is used for the first time, the user must:
1. Check the power supplies.
2. Check the patient circuit.
3. Perform a manual test.
4. Test the operation of the visual and audible alarms.
Each time the ventilator starts up and after every change in the circuit configuration, only the first
three stages are necessary.
PRESET
PRESET
PREVIOUS SETTING
PREVIOUS SETTING
ACV
ACV
SETTING
SETTING
TEST
TEST
Last test the 18/06/10
Double cirtcuit : PASS
NEW
NEW
SETTING
SETTING
Standby mode
Activation button on right side -->
FULL INTERNAL BATTERY IN CHARGE
Tuesday 13 February
Tuesday 13 February
19:32:45
19:32:45
Example of a screen showing the
internal battery charge indicator
Standby screen
Transition to standby takes
place after 60 seconds if the
touch screen is not pressed
Press the
button to
return to the previous screen
Summary of Contents for Elisee 350
Page 1: ...Elis e 350 Clinical Manual English Making quality of care easy Respiratory Care solutions...
Page 2: ...Making quality of care easy Respiratory Care solutions...
Page 4: ......
Page 6: ...iv...
Page 10: ...4 102 NOT013281 8...
Page 20: ...14 102 NOT013281 8...
Page 50: ...44 102 NOT013281 8...
Page 86: ...80 102 NOT013281 8...
Page 109: ...Making quality of care easy Respiratory Care solutions...