NOT013281-8
2.6 High-pressure oxygen port
The high-pressure oxygen port accepts oxygen at a pressure of between 240 kPa and 700 kPa.
When high-pressure oxygen is in use, it replaces ambient air as the ventilator’s air supply.
2.7 Low-pressure oxygen port
The low-pressure oxygen port can accept an oxygen supply at a pressure of up to 400 kPa.
2.8 Standard pneumatic nebuliser
The nebulisation function is available only if the Elisée 350 is supplied with high-pressure oxygen.
The driving gas is pure oxygen with a maximum pressure of 200 kPa and a flow rate of 20 L/min.
High-pressure O
2
port with high-
pressure O
2
hose inserted
Low-pressure oxygen port
Metal coupling for connecting an external
low-pressure oxygen supply
Nebuliser
Nebuliser tubing