XCell™ C410:V3 Controller
User Guide
27 repligen.com UG-C410v3-06
See
for details on Setting IP Addresses and Setting Connection.
Figure 12. Calibration screen
Analog Input Configuration/Calibration: Only accessible to the Administrator and Engineer Login. The
Analog Input Configuration/Calibration screen allows for the setup of analog inputs. For each analog
input the Administrator and Engineer will be able to configure the Engineering Units, the Minimum
Engineering Value, the Maximum Engineering Value, and perform a two-point calibration.
Note:
this should only be performed by a qualified metrology personal or Repligen Field Service
Engineer.
Table 9. Calibration screen display parameters
Field
Description
Engineering Units
Enter the Engineering Units for the selected analog input
Minimum Engineering
Value
Enter the Minimum Engineering value for the selected analog input
Maximum Engineering
Value
Enter the Maximum Engineering Value for the selected analog input
Calibration Point #1
Enter the first calibration point in Engineering Units in the range of Minimum
Engineering Unit to Maximum Engineering Unit
Calibration Point #2
Enter the second calibration point in Engineering Units in the range of
Minimum Engineering Unit to Maximum Engineering Unit