XCell™ C410:V3 Controller
User Guide
15 repligen.com UG-C410v3-06
the Siemens S7 - 1200 PLC. It is programmed using Siemens Step 7 Basic v13. The Operator Interface
Terminal (OIT or HMI) is a Siemens SIMATIC TP 1200 Comfort, programmed using Siemens WinCC
Advanced v13.
Figure 5. Electric Box (E-Box) connections
The Electronic enclosure includes the following connections:
j.
120/220vac power cable with plug
k.
Interconnect cable plug
l.
Operator Interface Terminal (OIT or HMI)
m.
Illuminated POWER ON/OFF selected switch
n.
Illuminated system STOP push button
o.
Alarm relay outputs (2)
p.
Harvest pump relay outputs (2)
q.
Ethernet plug
2.2
Primary pneumatic services
Air inlet (e) -
Located on the P-Box and provides an inlet to house compressed air source.
Recommended minimum air pressure requirement is ~50 psi / 3.4bar.
Somewhat higher inlet
pressures may be required, as needed, to generate higher flow rates or to drive pneumatic
instruments.
Exhaust/ vacuum line (f) -
Located on the P-Box.
This line is connected to a vacuum source. A
vacuum source should always be connected to the exhaust line to ensure reliable operation of the
WARNING:
Do not exceed 90 psi / 6.1 bar on the Air inlet.