Table 2. Utility requirements
Utility
Requirement
Additional considerations
Primary compressed air
Maximum 90 psig /
6.1 bar
Oil free, dry, filtered gas, i.e., medical grade air
Secondary pressure
Typically: 35 psi / 2.4bar,
Regulated oil free, dry, filtered air
Vacuum service
Minimum -12.5 psig
Vacuum supplied by a Repligen or customer
supplied local pump capable of maintaining ~-
12.5 psig with nominal flow as follows:
XCell ATF® 4 - 40L/min
XCell ATF® 6 - 60L/min
XCell ATF® 10 - 200L/min
Pump should be clean room compatible
Exhaust
Untreated discharge or user specified
Steam (SIP)
~30lbs/hr.
Applicable only to a steamable connection
between XCell ATF® Device and bioreactor vessel
Condensate drain
For SIP condensate drainage
Altitude
2000 m
This product has been assessed for a maximum
altitude of 2000 m.
Pollution degree
Pollution degree 2
This product is intended to be used in an
environmental of pollution degree 2. Pollution
degree 2 area is normally where only
nonconductive pollution can be present.
Temporary conductivity that is caused by
condensation is to be expected.
Current
0.6 AMP
Frequency
50/60Hz
Supply voltage
100 - 240VAC (+10%, -10%) Main supply voltage fluct/- 10%
Environment
temperature humidity
Ambient Temperature (0-
50° C) Low Humidity (0-
80% RH) Indoor (Dry)
Environmental specifications only apply to
controller. Controller is designed to be used
indoor (dry) and low humidity location