5.1
Fault-finding
Symptom
Cause
Action
No LEDs lit on RMI-QE.
Overvoltage, undervoltage or no
power.
Check voltage supply.
Damaged cable.
Check wiring.
RMI-QE status LEDs do
not correspond to radio
“PROBE STATUS” LEDs.
Radio link failure – radio probe
out of RMI-QE range.
Check position of RMI-QE, refer
to the operating envelope in the
relevant radio probe installation
guide. For further information, see
Section 6,
“Parts list”
.
Radio probe has been enclosed/
shielded by metal.
Review installation.
Radio probe and RMI-QE are not
partnered.
Partner radio probe and RMI-QE.
RMI-QE “ERROR”
LED illuminated. For
more information, see
page
2.4, “ERROR”
LED.
Radio probe and RMI-QE are not
partnered.
Partner radio probe and RMI-QE.
Exhausted radio probe batteries.
Change radio probe batteries.
Probe not switched on.
Check configuration and alter as
required.
Probe out of range.
Check position of RMI-QE, refer
to the operating envelope in the
relevant radio probe installation
guide. For further information, see
Section 6,
“Parts list”
.
Probe selection error.
Verify that one radio probe is
working and is correctly selected.
All four lower LEDs
flashing.
Wiring fault.
Check wiring.
Output overcurrent.
Check wiring, turn power to
RMI-QE off and on again to reset.
Summary of Contents for RMI-QE
Page 1: ...RMI QE radio machine interface Installation guide H 6551 8520 01 A...
Page 6: ...RMI QE installation guide iv Contents This page is intentionally left blank...
Page 26: ...RMI QE installation guide 2 16 RMI QE basics This page is intentionally left blank...
Page 44: ...RMI QE installation guide 3 18 System installation This page is intentionally left blank...
Page 48: ...RMI QE installation guide 4 4 Maintenance This page is intentionally left blank...
Page 53: ......