EN
4.5 Connecting the Electrical Devices
⇒
Connect the electronic device to the appliance socket (8) on
the back�
When connecting electrical appliances to the suction unit,
ensure that the maximum permissible power for connected
devices is not exceeded (see chapter 8, Technical data).
4.6 External Exhaust Air Duct
The exhaust air can be transported out of the laboratory via an external exhaust air duct (see accesso-
ries)�
The installation instructions are included in the external exhaust air duct�
When using suction devices in conjunction with an external exhaust air duct, a large amount of air
is extracted from the room per hour.
This can lead to a negative pressure, whereby, when using air-dependent fireplaces with gas,
liquid or solid fuels, poisonous gases (e.g. carbon monoxide) will be sucked into the (working)
space.
Depending on the structural situation, it is therefore essential to ensure there is an addition-
al air supply or negative pressure monitoring which is checked by the appropriate institutions
(e.g. chimney sweep or heating engineer).
5 Operation
The suction is controlled by the keys on the control panel (Fig� 2)�
5.1 Switching On
The suction unit is switched on and off via the on / off switch (3)�
♦ The suction unit performs an automatic filter cleaning sequence (for ap
-
prox� 8 seconds, loud vibrating noise)�
(see chapter 5�4)�
Afterwards, the suction unit is in the last set mode�
5.2 Select operating mode: Automatic operation / continuous operation
The extraction unit has two operating modes�
The set operating mode is shown in the display (20) / (22)�
• Automatic mode (20):
The suction is dependent on the electrical device connected to the appliance
outlet (8, Fig� 1)�
• Permanent operation (22):
Suction running continuously�
⇒
Press operating mode key (21)�
♦
Changing the operting mode�
To start or stop the unit during permanent operation, use the operating mode key (21). Do not use
the on / off switch for this.
If the suction to a connected electrical appliance (e.g. a handpiece) does not react as required
(e.g. does not activate, even though a connected device is in operation) then the switch-on thresh-
old for automatic operation must be adjusted (see chapter 5.5.3).
Fig. 4
8
Fig. 5
3
Fig. 6
21
20
22
- 9 -
Summary of Contents for SILENT TC
Page 1: ...AR KO JA ZH PL RU TR PT ES IT FR EN DE Made in Germany SILENT TC 21 6812 26112019 ...
Page 2: ......
Page 4: ... A ...
Page 6: ...C Wand wall mural parete pared стене min 50 mm min 2 inch ...
Page 7: ...DE ORIGINALBEDIENUNGSANLEITUNG Made in Germany SILENT TC 21 6812 26112019 ...
Page 43: ...FR TRADUCTION DU MODE D EMPLOI D ORIGINE Made in Germany SILENT TC 21 6812 26112019 ...
Page 61: ...IT TRADUZIONE DEL MANUALE DI ISTRUZIONI ORIGINALE Made in Germany SILENT TC 21 6812 26112019 ...
Page 79: ...ES TRADUCCIÓN DEL MANUAL DE INSTRUCCIONES ORIGINAL Made in Germany SILENT TC 21 6812 26112019 ...
Page 97: ...PT TRADUÇÃO DO MANUAL DE UTILIZAÇÃO ORIGINAL Made in Germany SILENT TC 21 6812 26112019 ...
Page 115: ...TR ORIJINAL KULLANIM KILAVUZUNUN ÇEVIRISI Made in Germany SILENT TC 21 6812 26112019 ...
Page 133: ...RU ПЕРЕВОД ОРИГИНАЛА ИНСТРУКЦИИ ПО ЭКСПЛУАТАЦИИ Made in Germany SILENT TC 21 6812 26112019 ...
Page 151: ...PL TŁUMACZENIE ORYGINALNEJ INSTRUKCJI OBSŁUGI Made in Germany SILENT TC 21 6812 26112019 ...
Page 169: ...ZH 此为原本使用说明书的翻译版本 Made in Germany SILENT TC 21 6812 26112019 ...
Page 185: ...JA 原文の取扱説明書の翻訳 Made in Germany SILENT TC 21 6812 26112019 ...
Page 201: ...KO 원본 사용 설명서의 번역본 Made in Germany SILENT TC 21 6812 26112019 ...
Page 219: ...AR اﻷصلية االستخدام تعليمات ترجمة Made in Germany SILENT TC 21 6812 26112019 ...
Page 240: ...2 x Montage assembly montage montaggio montaje сборка 7 8 9 2 2 ...
Page 241: ......