background image

DCO-0599 

Effective Date: 14-DEC-2018 

 

LBL-6025 Rev B 

Page  1 of 8 

 

 

  

Tricera Handpiece (R-10002)  

Instructions for Use 

PREFACE 

This manual provides the information to operate and maintain the 

Tricera Handpiece. It is essential to read and understand all the 

information in this manual before using or maintaining this device. 

Refer t

www.religncorp.com

 for the latest revision of this document. 

CONTENTS 

Preface ................................................................................................... 1 
Contents ................................................................................................. 1 
1. 

Symbols Used on Labeling ......................................................... 1 

2. 

Device Description ...................................................................... 1 

3. 

Indications for Use ...................................................................... 1 

4. 

Contraindications ........................................................................ 1 

5. 

Warnings and Cautions .............................................................. 1 

6. 

CompatIbility ............................................................................... 2 

7. 

How Supplied .............................................................................. 2 

8. 

Required Materials ...................................................................... 2 

9. 

Inspection, Maintenance and Testing ......................................... 2 

10.  Reprocessing Instructions .......................................................... 2 
11.  Storage ....................................................................................... 5 
12.  Limited Warranty ......................................................................... 5 
13.  Customer Service/Technical Support ......................................... 6 
Appendix A: Tricera Handpiece Components ....................................... 7 
Appendix B: Troubleshooting ................................................................. 8 
 

1. 

SYMBOLS USED ON LABELING  

 

Catalog Number 

 

Consult Instructions for Use 

 

Serial Number 

 

Non-Sterile 

 

Store in a cool dry 

place 

 

MRI unsafe 

 

Date of Manufacture 

 

Manufacturer

 

 

Type BF Applied 

Part 

 

Waste Electrical and 

Electronic Equipment. Must 

follow local law for 

equipment disposal. 

2. 

DEVICE DESCRIPTION 

The Tricera Handpiece is a reusable, electrically powered, hand-held 

surgical instrument that uses interchangeable shaver blades/RF 

probes to abrade and remove bone and accomplish the debridement 

of soft tissue.  The handpiece features a quick-locking mechanism for 

easy blade insertion and removal, and a suction irrigation port for 

removing irrigant from the operating site. 
The Tricera Handpiece is for use with the Relign Tricera System.  

3. 

INDICATIONS FOR USE 

The Relign Arthroscopic System is indicated for use in orthopedic and 

arthroscopic procedures.  The Fluid Management System of the Relign 

Arthroscopic System provides fluid distension and irrigation of the knee, 

shoulder, ankle, elbow, wrist and hip, and fluid suction during 

diagnostic and operative arthroscopic procedures.  The Shaver 

Blade/RF Probe of the Relign Arthroscopic System provides abrasion, 

resection, debridement, and removal of bone through its shaver blade; 

removal, ablation, and coagulation of soft tissue; as well as hemostasis 

of blood vessels through its shaver blade and probe.  Examples of uses 

of the product include resection of torn knee cartilage, subacromial 

decompression, and resection of synovial tissue in other joints. 

Contraindications:

 The electrosurgical probe should not be used in 

procedures where a nonconductive irrigant is used or with patients 

having cardiac pacemakers or other electronic implants. 

4. 

CONTRAINDICATIONS 

The Relign Arthroscopic System contains a large amount of metal 

components. Therefore, it is MRI unsafe. DO NOT use the Relign 

Arthroscopic System in conjunction with MRI, CT or RFID. 

5. 

WARNINGS AND CAUTIONS  

Please read this manual and follow its instructions carefully.  The words 

warning

caution

, and 

note

 carry special meanings and should be 

carefully reviewed: 

Warning

  The personal safety of the patient or physician may be 

involved.  Disregarding the information could result in injury 

to the patient or physician. 

Caution 

Special service procedures or precautions must be 

followed to avoid damaging the instrument.

 

Note 

Special information to make maintenance easier or 

important information more clear. 

 

An exclamation mark within a triangle is intended to 

alert the user to the presence of important operating 

and maintenance instructions in the literature 

accompanying the product.

 

Warning 

 

To avoid potential serious injury to the user and the 

patient and/or damage to this device, please note the 

following warnings: 

1)

 

  Federal law (United States of America) restricts this 

device to use by, or on order of, a physician. 

2)

 

  To avoid potential serious injury to the user and patient 

and/or damage to this device, read this operating 

manual thoroughly and be familiar with its contents prior 

to using this equipment. 

3)

 

  The Tricera Handpiece is designed to be used by a 

qualified physician familiar with arthroscopic 

procedures. It is the responsibility of the user to be 

familiar with indications, contraindications, potential 

complications and risks associated with the arthroscopic 

procedures being performed.  Misusing the Tricera 

Handpiece may cause damage to both the patient and 

system components. 

4)

 

  Carefully unpack the unit and check if any damage 

occurred during shipment.  DO NOT use the instrument 

if damage is detected.  Please refer to the Customer 

Service/ Technical Support section in this manual. 

Summary of Contents for R-10002

Page 1: ...asis of blood vessels through its shaver blade and probe Examples of uses of the product include resection of torn knee cartilage subacromial decompression and resection of synovial tissue in other joints Contraindications The electrosurgical probe should not be used in procedures where a nonconductive irrigant is used or with patients having cardiac pacemakers or other electronic implants 4 CONTR...

Page 2: ...ssory or device be sure to follow the instructions provided with the accessory or device 7 HOW SUPPLIED Warning The Tricera Handpiece is supplied non sterile It must be cleaned and sterilized before the first use Prior to each subsequent use it must be cleaned and sterilized The shipping box contains 1 ea Relign Tricera Handpiece 1 ea Shaver Blade RF Probe Cavity Brush Gordon Brush PN 721012 or eq...

Page 3: ...ece from the other system components a Remove the shaver blade RF probe from the handpiece b Dispose of all used shaver blades RF probes according to hospital regulations c Disconnect the handpiece from the Relign Controller 2 Attach the soaking cap to the cable connector Caution Always install the soaking cap prior to cleaning or sterilization If moisture reaches the connection pins of the cable ...

Page 4: ...ny additional moisture from the handpiece 10 2 STERILIZATION INSTRUCTIONS Clean and sterilize the handpiece prior to the first use and every use thereafter Caution Never sterilize the handpiece that has not been cleaned The success of the sterilization process depends on the previous state of cleaning Caution Always install the soaking cap prior to cleaning or sterilization If moisture reaches the...

Page 5: ...ade in a sterile irrigant and activate suction while testing the handpiece 2 Run the handpiece through its various functions modes and settings to ensure it responds properly to commands Ensure that all parts move freely and that there are no loose or missing components Note any unusual sounds or vibrations and note operating speeds 3 Refer any calibration or operating difficulties to your Relign ...

Page 6: ... term original purchaser as used in this warranty means that person or organization and its employees if applicable to whom the Tricera Handpiece was sold by Relign This warranty may not be assigned or transferred in any manner In order to make a warranty claim the original customer must obtain a Return Materials Authorization Number RMA Number and return instructions from Relign s Customer Servic...

Page 7: ...era Handpiece 1 Shaver Blade RF Probe Cavity Brush Gordon Brush PN 721012 or equivalent 2 Fluid Channel Brush Key Surgical PN AT 12 197 or equivalent 3 Motor Shaft Brush Gordon Brush PN 499NAD ST or equivalent Figure A3 Cleaning Brushes 1 Sterilization Tray 2 Sterilization Tray Lid 3 Rubber Holders Figure A4 Optional Sterilization Tray if available 1 2 3 1 Shaver Blade Cavity 2 Motor shaft 3 Fluid...

Page 8: ...The shaver blade does not turn The shaver blade is not inserted properly Remove the blade and reinsert it according to the Relign Tricera System instructions for directions on use of the Tricera Handpiece as part of the Relign Tricera System The handpiece will not run The handpiece is not properly connected to the controller Remove and reconnect the handpiece cable to the controller according to t...