Rev G
REF TEK 130 Multi-Channel Users Guide
5/14/2021
96223-00-UG
24
Reftek Systems Inc.
You may set the following parameters for each datastream:
o
Which input channel or channels to include
o
Sample rate
o
Data format (recording resolution)
o
Trigger type
The possible values for these parameters, as well as the possible combinations of parameters,
are different for the 130-MC with different hardware configurations, different versions of control
firmware, or both. For additional explanation of your 130-MC specific capabilities and
limitations, refer to the REF TEK
130 Command Line - Theory of Operations
manual.
Example configuration
The following steps illustrate how to write an example of a parameter file that sets up a 130-MC
with a vote trigger:
1.
Create the file in Notepad or a similar editor.
2.
Save the file as
trigger.prm
3.
Put the file on the disk
4.
Load the parameter file into the 130-MC using the following command:
{LP,0,\DSK1\trigger.prm
{PE,0
Erase the current copy of parameters from 92CB
{PS,0,92,test1,field station,1,12345,12345 at Chico
{PC,0,1,Vertical,A,I,120,300,180,degree,degree,1,131A,254,site4
{PC,0,2,Horiz,A,I,120,300,180,degree,degree,1,131A,254,site4
{PC,0,3,Horiz,A,I,120,300,180,degree,degree,1,131A,254,site4
{PD,0,RC,D,CO,200,1.2.3,120,T
Activate
Channels
Define
{PD,0,TG,10,12,0.1,6,6,10,30
{PD,0,TC,1,4,100,2,50
{PD,0,TC,2,4,50,2,30
{PD,0,TC,3,4,50,2,30
Datastream
Parameters
Record to Disk
CO=compressed
200 sps
1,2,3 channels included
Record length=120 sec
T=Vote trigger
10 second trigger window
12 Hz Low Pass trigger filter
0.1 Hz High Pass trigger filter
6=Total Trigger Votes
6=Total Detrigger Votes
10 second Pre-trigger
30 second Post-trigger
Channel 1
4 Trigger votes
level=100mG
2 Detrigger votes
level=50mG
General
Trigger
Channel
Trigger
Channel 2
4 Trigger votes
level=50mG
2 Detrigger votes
level=30mG
Channel 3
4 Trigger votes
level=50mG
2 Detrigger votes
level=30mG
Set informational
Station Parameters
1-3
{AL,0,EV,Y
{PM,0,D1,ATDT2143530609
Enable alert
Number to dial
callout
an event callout