89
5.3.2
Sensor Test Signals
Sensors vary in the number and type of signals they support. The following
table shows some common sensors used with the Ref Tek 130-01:
Sensor
Mass
Center
CalSig
CalEn
Notes
Outputs
Inputs
Inputs
Inputs
131A
-
-
-
1 ( +5v )
Self-test output while enabled
LE3Dlite
-
-
-
1 (GND)
Fixed self-test starts on falling
edge
L4C
- - 1 -
S13
- - 1 -
CMG-3ESP
3
1 (GND)
1
3 (GND)
Center enable > 0.2 S
CMG-3T
3
1 (GND)
1
3 (GND)
Center enable > 0.2 S
CMG-40T
3 - 1
3
(GND)
STS
3
1 ( +5v )
3
1 ( +5v )
Center enable > 0.5 S
5.3.3
Auxiliary Channels
Three auxiliary channels are provided on each sensor connector. Each
channel connects to a 12-bit A/D converter with an input range of +/- 10v.
These channels are read once per second and stored as left-justified, 2’s-
compliment, 16-bit values. This equates to a bit-weight of 305.2 uV. These
channels can be used to measure mass position or other auxiliary input data,
such as temperature and wind speed. The current readings of the auxiliary
channels are available on demand using the Auxiliary Data status response
(SS:AD).
Mass re-center
Mass re-centering can be performed on-demand. The Sensor Calibration –
Mass re-center (SK:M) command forces activation of the Center pins of a
specified sensor for 1-2 seconds. Automatic re-centering is enabled using the
PQ parameters. This assumes that the auxiliary input channels are
connected to the mass position outputs of a sensor.
Summary of Contents for 130-01
Page 16: ...130 01 Operations Refraction Technology Inc 2 ...
Page 54: ......
Page 82: ...130 01 Operations Refraction Technology Inc 68 Figure 51 PTO Connector ...
Page 87: ...Doc 1301 Ops Rev 2 9 1 10 30 08 Refraction Technology Inc 73 ...
Page 88: ...130 01 Operations Refraction Technology Inc 74 ...
Page 89: ...Doc 1301 Ops Rev 2 9 1 10 30 08 Refraction Technology Inc 75 ...
Page 96: ......