Page 1: ......

Page 2: ...th America 61 0 8 9445 4020 1 705 235 2169 Reflex South Africa Reflex South America 27 0 11 792 045 56 2 247 9504 Reflex Europe 44 0 1273 475 928 www reflexinstruments com Copyright 2008 Reflex Instru...

Page 3: ...en will appear Highlight Single Shot and press OK 4 You will now be asked if you would like to create a New Survey or add to an Existing Survey file To create a New Survey file named after the Hole ID...

Page 4: ...E and press OK 7 Enter the survey depth e g 100m When finished highlight CONTINUE and press OK Note This existing survey file named after the Hole ID Number can be used for all surveys performed in th...

Page 5: ...e information is incorrect highlight BACK and press OK Keep going back until you reach the information that is incorrect 9 Pull back 6 meters Assemble the instrument and lower through the bit face Ens...

Page 6: ...taken highlight REDO and press OK Repeat step 10 12 Pull the instrument to surface Undo the instrument from the top sub to expose the IR port Highlight DOWNLOAD point the IR ports at each other and pr...

Page 7: ...are more than one survey using the same Hole ID Number use the left and right arrows to scroll to the other survey depths in this file 14 To return to the MAIN menu scroll down and highlight MAIN MENU...

Page 8: ...appear Highlight Survey and press OK 3 The SURVEY menu screen will appear Highlight Multi Shot and press OK 4 A default survey file name will now appear Highlight NO and press OK 6 5 Now enter your s...

Page 9: ...en finished highlight CONTINUE and press OK 7 Enter the survey interval e g 9m or how often you would like to take a survey shot 8 You will now be asked if you would like to survey into the hole or ou...

Page 10: ...he survey instrument Press OK There is no need to set a survey hold off time b If the information is incorrect highlight BACK and press OK Keep going back until you reach the information that is incor...

Page 11: ...interval e g 9m MEASURE is highlighted press OK to take the shot Repeat this step until the drilling barrel is at the surface Note The shot can be taken while the drill rod is being taken off Note If...

Page 12: ...to end the survey 14 Pull the instrument to surface Undo the instrument from the top sub to expose the IR port Highlight DOWNLOAD point the IR ports at each other and press OK This will retrieve the f...

Page 13: ...15 Using the left and right arrows you can scroll through other survey depths 16 To return to the MAIN menu scroll down and highlight MAIN MENU Press OK 11...

Page 14: ...ight Survey and press OK 3 The SURVEY menu screen will appear Highlight Orientation and press OK 4 A default survey file name will now appear Highlight NO and press OK 12 5 Now enter your survey file...

Page 15: ...OOLFACE if orientating angled holes and press OK Note The Stored Zeroing value will be replaced by a new value if GRAVITY or MAGNETIC TOOLFACE is selected 7 Insert the instrument that is connected to...

Page 16: ...ero is displayed go to step 11 below Use Stored Zeroing Note only use the Stored Zeroing value if you are sure that it can be used for this survey 9 a If you are sure that the Stored Zeroing value can...

Page 17: ...plicable for the present survey highlight CONTINUE and press O K Go to step 11 below b If the Stored Zeroing is not applicable for the present survey highlight BACK and press OK Go to step 6 above and...

Page 18: ...for the orientation survey Note REFLEX EZ COM will hibernate if not used for 2 minutes Press the ON button to continue 12 Highlight INIT point the IR ports at each other and press OK 13 Run the instr...

Page 19: ...whether or not a good survey has been taken highlight REDO and press OK Repeat step 13 15 Retrieve the instrument Check that the Muleshoe lead has been stung Undo the instrument from the top sub to e...

Page 20: ...If the reading is not as desired highlight Enter desired Toolface and press OK Go to step 18 below b To save the results to file highlight End survey and Save results and press OK The Zeroing value i...

Page 21: ...highlight CONTINUE and press OK 19 a Rotate the rods clockwise by the amount displayed b To verify the toolface by a new measurement highlight RE MEASURE and press OK Repeat step from step 12 c To re...

Page 22: ...ravity values are displayed b To return to the MAIN menu highlight MAIN MENU and press OK c To view the stored Zeroing value highlight ZEROING and press OK 21 The stored Zeroing value is displayed BAC...