68
Maintenance
There is no regular maintenance service required for the MOTOmed.
Technical safety checks that comply with the medical product operator
regulation (Medical Devices Act) must be carried out at least every two
years. Please pay attention to follow the most recent version of this
regulation DIN EN 62353 VDE 0751,-1:2008.
If the MOTOmed is used for different patients within the provision
system of the health insurance, the visual inspection according to
chapter 12 has to be carried out. Worn parts (i.e. foot shell foam, han-
dles, expanders) should be replaced.
Recycling
The MOTOmed has a high quality all-metal construction: it is long
lasting, environmentally compatible and recyclable. Users are required
to dispose of such used electrical and electronic devices in accordance
to EC regulations 2002/96/EG-WEEE (Waster Electrical and Electronic
Equipment).
Please contact your MOTOmed representative for any questions.
page 87
page 78
CLEANING, MAINTENANCE, RECYCLING