background image

USER MANUAL
Thank you for purchasing the RCBS Digital Pocket Scale. 

Please read all operating instructions carefully prior to 

use. This electronic scale is a precision instrument, and will 

provide years of reliable service. Never load the scale with 

more than the maximum load capacity of 100 grams or 1500 

grains. Overloading will permanently damage the scale. 

Avoid any exposure to extreme heat or cold, your scale works 

better when operated at normal room temperature. Always 

allow the scale to acclimate to normal room temperature and 

allow for sufficient warm up time prior to calibration. This is 

done to ensure the internal components have stabilized and 

reached thermal equilibrium. 30 minutes will be a sufficient 

warm up period. 
Keep your scale in a clean environment, free of dust, dirt and 

moisture. Be aware of vibration and air currents as this can 

also affect the scales performance. Maintain an adequate 

distance from other electronic devices as some electronics 

can cause an electrical disturbance which can also affect 

the scales accuracy. Handle with care and apply all items to 

be weighed directly onto the platform or in the provided 

scale pan. This scale is built for its durability; however, try to 

avoid unnecessary rough handling as this may permanently 

damage the load cell. 
Avoid shaking, dropping or otherwise shocking the scale. This 

is a precision instrument and must be handled with extreme 

care.

 

ONLY OPERATE THE SCALE ON A STABLE, VIBRATION FREE 

SURFACE

Prior to use, turn on scale by pressing the “ON/OFF” key. 

Place the calibration weight onto the platform and allow 

unit to warm up for 30 minutes prior to use.

CALIBRATION

1.  Place scale on a flat surface.

2.  Press “ON/OFF” key to turn the scale on.

3.  When “0.0” or “0.000” is displayed, press and hold the 

“MODE” key until “CAL” is displayed

4.  Press the “MODE” key again, “50.00g” will flash.

5.  Place the 50 gram Calibration weight onto the center of 

the platform.

6.  Leave the calibration weight on the platform until “PASS” 

is displayed.

7.  Remove calibration weight.

8.  Scale will return to zero.

OPERATION

1.  Use scale only after successful completion of calibration.

2.  Use the “MODE” key to select the desired units of 

measure; Grams (g) or Grains (gn).

3.  Place the scale pan onto the platform and press the 

“TARE” key. This will tare out the scale pan, so objects 

can be placed inside of the scale pan for weighing.

4.  If the scale pan is not being used, press the “TARE” with 

nothing on the platform, this will allow for placing 

objects directly onto the platen for weighing. 

It is good practice to periodically check the scale by placing 

the calibration weight onto the platen. If the displayed weight 

is off by more than +/- .01 g or .1 gn. ; re-calibrate the scale.