MagCore HF48 Service Manual
Revised: February,2014
Ver. 2014-2
Write in home distance
1.
Please refer to “installation” → ”Test home value” → ” Measure the interval
between Y axis and costal plate”→” Measure the level height of X-axis and costal
plates at area A and area B” for the information about the previous steps to make
sure whether accuracy of the machine is shifted before inputting the home distance.
2. Prepare the 4 values (Y-a/X-a/Y-b/X-b/) measured by the previous step-“Test
home distance”.
3. Turn on the power.
4. I n p u t t h e va l u e s Y - a i n to t h e Y t a b 's ( 4 a )
column → (4b) press extension → (4c) press write in.
input
input
write in
press extension
press extension
press extension
write in
4a
5a
6c
4b
5b
6b
4c
5. I n p u t t h e v a l u e s X - a i n t o t h e X t a b 's ( 5 a )
column → (5b) press extension.
6. I n p u t t h e v a l u e s X - b i n t o t h e X t a b 's ( 6 a )
column → (6b) press extension → (6c) press write in.
input
6a