MagCore HF48 Service Manual
Revised: February,2014
Ver. 2014-2
Installation
Please make sure that the voltage of the
outlet is between 220V and 240V. Keep
your hands dry. Please also make sure
that the machine is turned off when you
insert the plug.
Please note the weight of the top cover of
part chamber. Two people are required to
lift it up.
Place the system in a suitable location and use the leveling glides to
fine-tune the elevation. Remember to tighten the screws.
Cut the fixation belt and close the top cover of part chamber.
Cut the fixation belts on the racks.
Plug in the power cord and turn on the main power.
Turn on the power and the system will start initializing setup
automatically.
Please set the local time (refer to page 24) for
the first installation, and perform the “Piston
Test” (page 20).
Please finish reading the operation manual
before operating.
Rotate and pull out the knobs on both sides of the top cover of part
chamber using coins.
Open the top cover of part chamber.
1
2
!
Next Step
Next Step
Next Step
Next Step
Next Step