Profinet
MI3
99
17.2.2.1
Station Parameters
The parameters of the station substitute are used to set the overall settings of the Profinet IO node.
Some of the setting are used in the modules as default settings and can be optionally overwritten
within the module configuration.
Parameter
Description
Setting
Temperature unit
Set the temperature unit
Celsius
Fahrenheit
Message diagnostics
The diagnostics information of all modules are not
transferred to the Profinet I/O controller
message inactive
transferred to the Profinet I/O controller
diagnostic alarm
Value behavior on module fault
All process data are immediately reset
set process data to zero
All process data maintain the last status
set process data to last value
Bottom temperature of output 1
Set Bottom temperature of analog output 1
0…3000°C /°F
Top temperature of output 1
Set Top temperature of analog output 1
0…3000°C /°F
Bottom temperature of output 2
Set Bottom temperature of analog output 2
0…3000°C /°F
Top temperature of output 2
Set Top temperature of analog output 2
0…3000°C /°F
Source (head) for output 1
Set head number
1..8
Source (type) for output 1
Set type of source for output 1
Internal temperature
Object temperature
Analog output 1 mode
Set output mode
TCJ
TCK
TCR
TCS
0…5V
0…10V
tristate (disabled)
Source (head) for output 2
Set head number
1..8
Source (type) for output 2
Set type of source for output 2
Internal temperature
Object temperature
Analog output 2 mode
Set output mode
0…20mA
4…20mA
0…5V
0…10V
tristate (disabled)