Instructions for Use MPR1 DATALOGGER
Manufacturer:
RAUMEDIC AG, Hermann-Staudinger-Strasse 2, 95233 Helmbrechts, Germany
Page 36 of 98
Software version
Confirm the selected Function with OK. This
function is only available if data are stored on the
device.
The trend data of the last measurement are
displayed on the device. Display of measurement
curves (data from SP memory mode) is not
possible.
The function "View last measurement" is indicated
by the reverse highlighted flashing symbol
the status line. In addition, the alarm LED flashes
yellow in this operating mode, so that the user
recognizes that the device is not in monitoring
mode.
4.3 Alarm limit selection
From here you carry out alarm limit selection. By pressing the Act key, the alarm limit settings of the last
measurement are adopted for the current measurement.
The alarm limits can be set over the entire measurement range. Please note that thresholds set
outside the physiological range can render the alarm system useless.
Use caution when selected the last alarm limits (Act key), as the previous user could have set
the thresholds to pointless values. Subsequently always check the actually set thresholds
(which are also the displayed values).
By pressing the Std key, the alarm limits from the operator settings are activated. Thus for constantly
repeating, similar, monitoring tasks, the time consuming setting of alarm limits is simplified.