Function
Description
Y Axis Log
Scale
Changes the Y axis scale of the
Plasmagram
view to a logarithmic scale.
X Axis Display
Units
Specifies the units used for the X axis in the
Plasmagram
view:
Time (ms)
or
Calibrated Units
.
Screen Saver Time
Specifies how long the device is idle before the screen saver turns on.
By default, if the device has been idle for 60 minutes, the screen saver turns on. The screen saver
can be disabled or set to turn on 5, 10, 20, or 60 minutes after the device has not been used.
Warn/Require
Calibration
Specifies the time interval to prompt a calibration warning.
By default, the calibration warning time interval is every 8 hours. The
Warn/Require Calibration
time
can be set for every 4, 8, 12, or 24 hours. Set to 0 to disable calibration warnings.
Audible Alarm
Specifies if an audible alarm should sound when a selected substance is detected.
Auto Save
Specifies the following:
• The type of automatic saving of data to be used:
Never
,
Every Alarm
, or
Every Sample
.
NOTE:
Only one of the three buttons can be selected.
• If the device should
Prompt for Name
for the file being saved and
Prompt for Notes
when
automatically saving data in the
Every Alarm
and
Every Sample
options.
NOTE:
Both boxes may be checked for these options.
NOTE:
Whenever the
Prompt for Name
or
Prompt for Notes
screen box displays,
you are given the option of using the built-in key panel to type in inform-
ation.
Set Defaults
Resets only the
Options
screen settings back to factory defaults.
3. Do one of the following:
•
To save all values using the touchscreen, press
OK
.
•
To save all values using the navigation button, press
Enter
.
Configuration
Accesses the
Configuration screen where analysis and sample times can be adjusted. This function is only
available to Rapiscan Systems.
Substances
Enables you to view the substances available for detection, as well as add, modify, and delete substances. You
cannot modify or delete substances from the custom library. You can only modify and delete substances that
50
Rapiscan Systems Proprietary and Confidential
CHAPTER 4 │ User's Guide
Summary of Contents for Mobile Trace
Page 1: ...MobileTrace User s Guide For P0007027 MA001133 Revision ...
Page 2: ......
Page 31: ...This page intentionally left blank to ensure new chapters start on right odd number pages ...
Page 60: ...CHAPTER 4 Menu Functions Main Screen 44 Menu Functions 45 ...
Page 164: ...CHAPTER 9 Parts and Accessories Device and Replacement Part Numbers 148 ...
Page 167: ...This page intentionally left blank to ensure new chapters start on right odd number pages ...
Page 178: ...For P0007027 MA001133 Revision C ...