NOTE:
For more dirty or dusty environments, you may need to clean the membrane more often.
5. Inspect the membrane for tears, punctures, incorrect alignment, accumulated dirt, or other damage. If
necessary, replace the membrane.
see "Replace the Membrane " on page 124)
Clean the Desorber or Vapor Plate
1. Using the Rapiscan Systems supplied canned air, blow air into the center hole on the back of the
desorber and blow air through the slot on the front of the desorber.
Figure 7-7
Blow dust and dirt from desorber
2. Using a Rapiscan Systems-supplied alcohol swab, gently clean the white seal disc and desorber or vapor
plate.
118
Rapiscan Systems Proprietary and Confidential
CHAPTER 7 │ User's Guide