Return to Operation
Make all entries in your device’s maintenance log as directed by your administrator.
Weekly Maintenance
To keep the device operating at peak performance, the procedures listed in this section should be performed
weekly in the step-by-step order listed. Also, you must verify shift maintenance has been performed.
CAUTION:
When maintaining the device, you must wear clean cotton gloves and use only Rapiscan
Systems-supplied consumables, parts, and supplies.
NOTE:
The ability to perform some maintenance procedures may depend upon the privilege levels
assigned.
Power Down the Device
1. Select Menu > Logoff.
2. Press the power button so the device is
OFF, and allow the device to cool for 5–10 minutes.
WARNING:
Device may be hot. You must let the device cool prior to opening the device.
Clean and Inspect the Membrane
1. Pull the two metal rings on either side of the cover to remove the desorber cover.
Figure 7-3
Remove desorber cover
2. Unscrew the retaining ring to remove the desorber or vapor plate and set aside.
116
Rapiscan Systems Proprietary and Confidential
CHAPTER 7 │ User's Guide