Command reference
R&S
®
ZNL/ZNLE
940
User Manual 1178.5966.02 ─ 19
Parameters:
<Boolean>
Enables (ON) or disables (OFF) the receiver power calibration
for the received waves b
<PhyPt>
.
*RST:
OFF
Example:
*RST; CORR:POW:ACQ BWAVe,1,PORT,2
Perform a receiver power calibration of the wave b1 using port 2
as a source port, assuming that the source power a
2
is correct.
A through connection from port 2 to port 1 is needed.
CORR:POW?
Check whether the calibration is applied (the response is
1
).
CALC:PAR:MEAS 'TRC1', 'B1D2'
Select b
1
as a measured quantity for the default trace.
CALC:MARK ON; MARK:Y?; :SOUR:POW?
Create marker no.
<Mk>
in the center of the sweep range and
query the measurement value. The calibrated power of the
received wave b
1
is approx. equal to the default source power
value.
See also example for
Manual operation:
See
[SENSe<Chn>:]CORRection:PSTate?
Gets the power calibration label of the active trace in channel <Chn>.
See
Chapter 9.5.6.3, "Power calibration labels"
on page 275 for the possible return
values.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Usage:
Query only
Manual operation:
See
[SENSe<Ch>:]CORRection[:STATe]
<Boolean>
Enables or disables the system error correction for channel
<Ch>
.
Suffix:
<Ch>
.
Calibrated channel number
Parameters:
<Boolean>
Enables (ON) or disables (OFF) the correction.
*RST:
ON
Example:
*RST; :CORR?
Reset the instrument and query whether channel 1 is system
error corrected. The response is
1
.
Manual operation:
See
VNA command reference