Command reference
R&S
®
ZNL/ZNLE
647
User Manual 1178.5966.02 ─ 19
Manual operation:
See
CALCulate:CALValidate:RESPonse:IMAGinary:LOWer
<Port>, <Value>
CALCulate:CALValidate:RESPonse:IMAGinary:UPPer
<Port>, <Value>
CALCulate:CALValidate:RESPonse:REAL:LOWer
<Port>, <Value>
CALCulate:CALValidate:RESPonse:REAL:UPPer
<Port>, <Value>
Defines the upper and lower cal validation limits for the
REAL
trace format (see
).
Use
CALCulate:CALValidate:CHARacteriza
to select the reference characteri-
zation.
Parameters:
<Port>
<Value>
*RST:
-200 (LOWer) | +200 (UPPer)
Default unit: mU
Manual operation:
See
CALCulate:CALValidate:RESPonse:MAGNitude:LOWer
<Port>, <Value>
CALCulate:CALValidate:RESPonse:MAGNitude:UPPer
<Port>, <Value>
Defines the magnitude cal validation limits for the
MLINear
trace format (see
).
Use
CALCulate:CALValidate:CHARacteriza
to select the reference characteri-
zation.
Parameters:
<Port>
Physical analyzer port no.
<Value>
*RST:
-0.2 (LOWer) | +0.2 (UPPer)
Default unit: dB
Manual operation:
See
CALCulate:CALValidate:RESPonse:PHASe:LOWer
<Port>, <Value>
CALCulate:CALValidate:RESPonse:PHASe:UPPer
<Port>, <Value>
Defines the phase cal validation limits for the
MLINear
trace format (see
).
Use
CALCulate:CALValidate:CHARacteriza
to select the reference characteri-
zation.
Parameters:
<Port>
<Value>
*RST:
-2 (LOWer) | +2 (UPPer)
Default unit: degrees
Manual operation:
See
VNA command reference