VNA GUI reference
R&S
®
ZNL/ZNLE
415
User Manual 1178.5966.02 ─ 19
Power Meter
Shows a list of all power meters that have been properly configured. See
The bordered label below displays the type and serial number of the selected power
meter.
Remote command:
CALCulate<Ch>:PARameter:MEASure
'TraceName', 'PmtrD1 | ...
CALCulate<Ch>:PARameter:SDEFine
'TraceName', 'PmtrD1 | ...
Auto Zero
Initiates an automatic zeroing procedure of the selected power meter.
The power meter must be disconnected from the RF power; see
on page 307. A message indicates that zeroing is finished.
Remote command:
SYSTem:COMMunicate:RDEVice:PMETer<Pmtr>:AZERo
Log Errors
Enables the transfer of error messages for external devices to the error log
C:\ProgramData\Rohde-Schwarz\ZNL-FPL\log\Error.log
.
Remote command:
n/a
Source Port
Selects one of the available test ports of the analyzer as a source of the stimulus sig-
nal.
Remote command:
CALCulate<Ch>:PARameter:MEASure
'TraceName', 'PmtrD1 | ...
CALCulate<Ch>:PARameter:SDEFine
'TraceName', 'PmtrD1 | ...
Show as
Selects the physical unit of the displayed trace. It is possible to display the measured
"Voltage" V or convert it to a "Power" according to the formula
P = V
2
/Re(Z
0
).
Z
0
denotes the reference impedance of the source port. The reference impedances are
defined in the "Balanced Ports" dialog (see
Chapter 10.7.1.3, "Balanced Ports dialog"
Remote command:
10.8
Format softtool
The "Format" softtool allows you to define how the measured data is presented in the
diagram area.
Access:
[Format]
Format softtool