background image

Instrument Tour

R&S

®

ZNH

22

Getting Started 1334.5979.02 ─ 03

Risk of instrument damage

To avoid damage to the tracking generator output, never apply currents
greater than 600 mA or voltages greater than 20 V to the BNC connectors if
the BNC connectors are not configured as BIAS output ports.
If they are configured as BIAS ports, never apply currents greater than 600
mA or voltages greater than 50 V.

Headphone jack

The 3.5 mm connector for headphones has an internal impedance of approxi-
mately 10 Ω.

USB type A connector

The USB port provides the USB interface to connect a memory stick and store
data sets or screenshots. It can also be used to control the operation of the exter-
nal power sensor.

Risk of electrical shock

To avoid electrical shock, the DC input voltage must never exceed the value
specified on the housing.

Risk of instrument damage

Make sure not to overload the RF input and keep within the maximum
allowed signal levels. Refer to the datasheet for the maximum allowed
signal levels.

To avoid damage to the coupling capacitor, input attenuator or the mixer,
the DC input voltage of 50 V must never be exceeded.

Top View

www.allice.de

Allice Messtechnik GmbH

Summary of Contents for ZNH

Page 1: ...R S ZNH Handheld Vector Network Analyzer Getting Started 1334597902 Version 03 R 2 www allice de Allice Messtechnik GmbH...

Page 2: ...18 R S ZNH26 1321 1611 76 equivalent to 1321 1611 26 2020 Rohde Schwarz GmbH Co KG M hldorfstr 15 81671 M nchen Germany Phone 49 89 41 29 0 Email info rohde schwarz com Internet www rohde schwarz com...

Page 3: ...ons and the product documentation in a safe place and pass them on to the subsequent users Riesgo de lesiones y da os en el instrumento El instrumento se debe usar de manera adecuada para prevenir des...

Page 4: ...tation gut auf und geben Sie diese an weitere Benutzer des Produkts weiter Risque de blessures et d endommagement de l appareil L appareil doit tre utilis conform ment aux prescriptions afin d viter l...

Page 5: ...the device manual or data sheet for the exact designation of the charger If the batteries are improperly charged there is a risk of explosion which can cause serious personal injury 5 The charging te...

Page 6: ...nsulte en el manual o en las especificaciones t cnicas del equipo la denominaci n exacta del cargador Si las bater as se cargan de forma incorrecta existe peligro de explosi n lo que podr a causar les...

Page 7: ...dbuch oder Datenblatt des Ger tes f r die genaue Bezeichnung des Ladeger tes Wenn Batterien unsachgem geladen werden besteht Explosionsgefahr was zu schweren Verletzungen bei Personen f hren kann 5 Di...

Page 8: ...ns le manuel ou la fiche technique de l appareil Une recharge incorrecte des batteries peut entra ner des explosions susceptibles de causer des blessures graves 5 Recharger imp rativement des temp rat...

Page 9: ...qualified engineers provides support and works with you to find a solution to your query on any aspect of the operation programming or applications of Rohde Schwarz products Contact information Conta...

Page 10: ...ZNH 7 3 Preparing for Use 8 3 1 Putting into Operation 8 3 1 1 Unpacking and Checking the Instrument 9 3 1 2 Accessory List 10 3 1 3 Setting up the R S ZNH 10 3 1 4 Using the AC Adapter 12 3 1 5 Batte...

Page 11: ...Contents R S ZNH 4 Getting Started 1334 5979 02 03 www allice de Allice Messtechnik GmbH...

Page 12: ...programming examples and information on maintenance and instrument interfaces Includes the contents of the getting started manual The online version of the user manual provides the complete contents...

Page 13: ...device ID of your instrument which you can find on a label on the rear panel 1 4 Release Notes Open Source Acknowledgment The release notes list new features improvements and known issues of the cur r...

Page 14: ...t demanding customer requirements Offering touchscreen input the analyzer enhances user experience in making measurements fast and easy This user manual contains a description of the functionality tha...

Page 15: ...damage due to inappropriate operating conditions Specific operating conditions are required to ensure accurate measure ments and to avoid damage to the instrument Observe the information on appropriat...

Page 16: ...tromagnetic interference EMI may affect the measurement results To suppress generated electromagnetic interference EMI Use suitable shielded cables of high quality For example use double shielded RF a...

Page 17: ...1 2 Accessory List The instrument comes with the following accessories Power supply cable and adapter set Li ion rechargeable battery USB2 0 cable A Mini Side strap Getting Started printed manual Docu...

Page 18: ...support on the back of the R S ZNH Before you turn on the R S ZNH you should insert the lithium ion battery inclu ded in the delivery into the battery compartment located at the back of the R S ZNH In...

Page 19: ...the DC port on the left side of the R S ZNH item 1 of Figure 3 1 Make sure to fully insert the AC adapter plug into the DC port Depending on the system you need firmly connect the appropriate power c...

Page 20: ...battery icon indicated in the display screen is illustrated below Figure 3 2 Battery charging and discharging process Charging time is about three hours when the R S ZNH is in inactive mode i e R S ZN...

Page 21: ...ernal charger and sup ply it with power via the AC power adapter An amber LED on the charger indicates the charging process The LED turns to green when the battery is fully charged A red LED on the ch...

Page 22: ...als If you cannot discharge the battery it may be deep discharged Charge the battery for 0 5 hours and check again If you cannot charge the battery it may be overcharged Discharge the battery and chec...

Page 23: ...1 6 4 End of Life The capacity of the battery decreases after it has gone through numerous charge cycles and nearing its end of life When the battery is dead do not open the bat tery Do not dispose ba...

Page 24: ...ary of LED indication on POWER key LED indication on Power key Descriptions Green LED Instrument is in operation mode Blue LED Instrument is in switch off mode with a fully charged battery A blinking...

Page 25: ...4 1 Front view of R S ZNH 1 RF port 1 2 BNC connector 3 Headphone jack 4 USB ports 5 RF port 2 6 Touch sensitive screen area 7 Softkey labels on display 8 Softkey 9 Systems keys 10 DC port behind pro...

Page 26: ...ain substances that may damage the instrument For example cleaning agents that contain a solvent may damage the front panel labeling plastic parts or the display Never use cleaning agents such as solv...

Page 27: ...With a single RF port it is possible to generate a stimulus signal and measure the response signal in reflection For measurement examples see CAT mea surements in the R S ZNH user manual With two RF...

Page 28: ...e is selected in the SWEEP menu The trigger threshold is similar to that of TTL signals When the BNC connector is configured as reference input you can apply a 10 MHz external reference signal to it f...

Page 29: ...xi mately 10 USB type A connector The USB port provides the USB interface to connect a memory stick and store data sets or screenshots It can also be used to control the operation of the exter nal pow...

Page 30: ...can also use the DC connector to charge the battery Kensington lock slot A Kensington lock can be anchored to the R S ZNH housing to secure it to a workstation mechanically 4 4 Right View 1 Min USB p...

Page 31: ...he micro SD card protective cap to access to the micro SD card slot You can use the micro SD card to store datasets or screenshots 4 6 Display Overview The display area has a touch sensitive screen th...

Page 32: ...er touch the screen with ball point pens or other pointed objects with sharp edges It is recommended that you operate the touchscreen by finger only As an alternative you may use a stylus pen with a s...

Page 33: ...ement 13 Brochure 6 C Calibration certificate 6 D Data sheet 6 Documentation overview 5 E Electrostatic discharge 8 ESD 8 G Getting started 5 I Instrument tour 18 O Open source acknowledgment OSA 6 P...

Reviews: