Command Reference
R&S
®
ZNC
616
User Manual 1173.9557.02 ─ 13
Example:
CALC4:PAR:SDEF "Ch4Tr1", "A1"
Create channel 4 and a trace named "Ch4Tr1" to measure the
wave quantity a
1
. The trace automatically becomes the active
trace.
SENS4:SWE:SRCP 2
Select drive port 2 for the active trace.
CALC4:PAR:MEAS? "Ch4Tr1"
Query the measurement result for "Ch4Tr1". The response is
'A1D2SAM'
.
[SENSe<Ch>:]SWEep:STEP
<StepSize>
Sets the distance between two consecutive sweep points.
Note:
This setting is valid for sweep types with equidistant sweep points only. It does not
apply to logarithmic and segmented sweeps.
Suffix:
<Ch>
.
Channel number
Parameters:
<StepSize>
Stimulus step size. The minimum step size is equal to the sweep
span divided by the maximum number of points (
) minus one, the maximum step size is equal to
). See also
description of manual control and program example below.
Range:
Depending on the other channel settings; see above.
*RST:
Depending on the analyzer model. The default step
size is equal to the default sweep span of the ana-
lyzer divided by the default number of points minus
one.
Default unit: Hz
Example:
*RST; :SWE:STEP?
Query the default step size. For a 8.5 GHz analyzer, the response
is
42499500
. 42499500 Hz = (8.5 GHz - 100 kHz) / 200.
SWE:STEP UP
Increase the step size.
FREQ:STOP?; :SWE:POIN?
Query the stop frequency of the sweep and the number of points.
Increasing the step size has changed both values.
Manual operation:
See
[SENSe<Ch>:]SWEep:TIME
<SweepDuration>
Sets the duration of the sweep ("Sweep Time"). Setting a sweep time disables the auto-
matic calculation of the (minimum) sweep time; see
.
SCPI Command Reference