GUI Reference
R&S
®
ZNC
194
User Manual 1173.9557.02 ─ 13
Ref Mkr
Creates a reference marker and assigns it to the active trace. When a marker is created,
a triangle labeled "R" is positioned on the trace and the marker coordinates are displayed
in the info field.
The stimulus position of the active reference marker can be entered in the "Ref Marker
Stimulus" entry field. The default position is the start of the sweep range or the position
of the last active marker.
The reference marker defines the reference value for all markers that are in "Delta
Mode".
Remote command:
All Markers Off
Removes all markers from the active trace. Markers on other traces are not affected. The
removed markers remember their properties (stimulus value, format, delta mode, num-
ber) when they are restored. The marker properties are definitely lost when the associ-
ated trace is deleted.
Tip:
To delete a single marker, drag it into vertical direction to release it from the trace
and drop it onto the "Delete" icon.
Remote command:
Coupled Markers
Couples the markers of all traces in the active recall set to the markers of the active trace.
While marker coupling is active, the active trace markers assume the role of master
markers; the other markers behave as slave markers, following any change of position
of the master marker. Coupled markers allow you to compare different measurement
results (assigned to different traces) at the same stimulus value (i.e., at the same mea-
surement point).
See also
Remote command:
4.2.6.2
Marker > Marker Props
Modifies the properies of a marker created previously (see
, on page 192). The functions (except "Export Markers...") are unavailable if the
active trace contains no markers.
Access:
TRACE > MARKER key or Alt + Shift + G
Trace Settings