Concepts and features
R&S
®
ZNB/ZNBT
94
User Manual 1173.9163.02 ─ 62
4.1.4.3
Parallel measurements on multiple DUTs
The support for parallel measurements on multiple DUTs depends on the analyzer
type:
●
For a 2-port R&S
ZNB or a 4-port R&S
ZNB without internal second source (see
Chapter 4.7.14, "Internal second source"
not
available
●
For a 4-port R&S
sure one DUT, while port 3 and port 4 can measure another one (using the same
stimulus settings).
●
Each internal generator of a R&S
ZNBT can drive multiple VNA ports simultane-
ously. Because of this, even without
a single DUT or multiple DUTs can be measured in parallel (using the same stimu-
lus settings).
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
1
18
19
20
21
22
23
24
17
DUT1
DUT2
VNA
●
, to reduce "crosstalk" between the DUTs
a frequency offset can be applied between the corresponding port groups (see
"Parallel Measurements with Frequency Offset"
●
Parallel
Frequency Conversion Measurements
require identical conversion settings
for all port groups
●
Parallel
not
supported
●
Parallel measurements are
not
are used
Arbitrary Drive Port Order (R&S
ZNBT only)
When performing parallel measurements on multiple DUTs, by default the R&S
ZNBT
firmware automatically selects the ports that should be driven in parallel. However, due
to possible crosstalk between DUT ports, it can be beneficial to prevent some ports
Basic concepts