Command reference
R&S
®
ZNB/ZNBT
1351
User Manual 1173.9163.02 ─ 62
This setting applies to all physical ports and all channels.
Parameters:
<DeEmbNwType>
SENDed | GLOop | OFF | ALL | DIFFerential | PSET | RENorm |
OFFSet | AOFFset
Selects the offset or de-/embedding type to be used, if config-
ured.
ALL means that all available offset and de-/embedding informa-
tion is applied. The <DeEmbType> can be omitted in this case.
*RST:
OFF
<DeEmbType>
EMBedding | DEEMbedding
Selects between embedding and deembedding for the selected
<DeEmbNwType>.
Must be specified for network types that can be used for embed-
ding and deembedding
Manual operation:
See
7.3.15.4
SOURce:POWer:CORRection:TCOefficient...
These commands allow you
The "Power Meter Transmission Coefficients" dialog allows you to modify the results of
a scalar power calibration to account for additional two-port devices (with known trans-
mission characteristics) in the test setup.
SOURce<Ch>:POWer<PhyPt>:CORRection:TCOefficient:CALibration
<Boolean>
Selects the position of the additional two-port in the test setup.
Note
that it is now possible to have two-ports at both positions (see
CORRection:TCOefficient[:STATe]
). This command is kept for compatibility rea-
sons.
Suffix:
<Ch>
.
Calibrated channel number. This parameter is ignored; the trans-
mission coefficient settings are valid for all channels.
<PhyPt>
Calibrated port number. This parameter is ignored; the transmis-
sion coefficient settings are valid for all sources.
Parameters:
<Boolean>
ON
–
two port at power meter (during calibration)
OFF
–
two port at DUT (during measurement)
*RST:
ON (notice that the correction is switched off after a
reset)
SCPI command reference