Command reference
R&S
®
ZNB/ZNBT
1232
User Manual 1173.9163.02 ─ 62
[SENSe<Ch>:]CORRection:COLLect:DELete
[<CalName>]
Deletes system error correction data generated and stored previously.
Suffix:
<Ch>
.
Channel number of the calibrated channel
Setting parameters:
<CalName>
Name of the calibration (string parameter) defined together with
the calibration type (
[SENSe<Ch>:]CORRection:COLLect:
ALL
- the analyzer deletes all calibrations.
If nothing is specified the analyzer deletes the last system error
correction stored by means of
*RST:
-
Example:
CORR:COLL:METH:DEF 'Test',RSHort,1
Select a one-port normalization at port 1 with a short standard
as calibration type.
CORR:COLL:SEL SHOR,1
Measure a short standard connected to port 1 and store the
measurement results of this standard.
CORR:COLL:SAVE:SEL
CORR:COLL:DEL ALL
Calculate the system error correction data and apply it to the
active channel, then delete the data.
Usage:
Setting only
Manual operation:
See
[SENSe<Ch>:]CORRection:COLLect:DETector
<CalName>[, <CalDetectorType>]
Selects the detector type during manual calibration <CalName>.
Suffix:
<Ch>
.
Channel number
Parameters:
<CalName>
Name of a manual calibration previously created e.g. using
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine
<CalDetectorType>
NORMal | AVERage
AVERage
Average detector, recommended for channels measuring traces
using detectors other than the NORMal one
NORMal
Normal (sample) detector, recommended for all other applica-
tions
*RST:
NORMal
SCPI command reference