R&S UPZ
Manual Operation
Link Establishment using an R&S UPV
Operating Manual 1120.8610.14 - 0
5
40
Table 2-1 List of settings and IEC/IEEE-bus commands: R&S UPL
Setting
Effect
IEC/IEEE-bus command
Switcher OFF
ON
R&S UPZ control is switched on
SWITcher:STATe ON|OFF
Input
A
n
Channel n of input switcher is switched to
busbar A
SWITcher:INPA <n>
Input
B
n
Channel n of input switcher is switched to
busbar B
SWITcher:INPB <n>
Output
A
n
Channel n of output switcher is switched to
busbar A
SWITcher:OUTA <n>
Output
B
n
Channel n of output switcher is switched to
busbar B
SWITcher:OUTB <n>
Tracking
OFF
B vs A
OUT vs INP
ALL
An additional selection window is opened
for channel tracking.
without tracking
only busbar B versus A
only output versus input
both tracking modes simultaneously
SWITcher:TRACking OFF
SWITcher:TRACking BVSA
SWITcher:TRACking OVSI
SWITcher:TRACking ALL
B vs A
n
The channel on busbar B is offset by <n>
versus A
SWITcher:OFFSet:BVSA <n>
Out vs Inp
n
The channel of the output switcher is offset
by <n> versus the input switcher
SWITcher:OFFSet:OVSI <n>
2.6 Link Establishment using an R&S UPV
The following steps illustrate how to establish a connection between an R&S UPZ and
an R&S UPV.
If the R&S UPZ needs to be controlled via RS232, open the Config Panel from the
menu bar:
Fig. 2-8 Config Panel: R&S UPV
The selection “Used for“ needs to be set to “Switcher“. If the COM interface was used
for remote control previously, the R&S UPV must be rebooted after the change.