Remote Control Commands
R&S
®
SMJ100A
764
Operating Manual 1403.7458.32 ─ 14
Parameters:
<State>
ON | OFF | AUTO
ON
Internal level control is permanently activated.
OFF
Internal level control is deactivated; Sample & Hold mode is acti-
vated.
AUTO
Internal level control is activated/deactivated automatically
depending on the operating state.
*RST:
AUTO
Example:
POW:ALC ON
activates automatic level control for RF output A.
Manual operation:
See
[:SOURce<hw>]:POWer:ATTenuation:RFOFf:MODE
<Mode>
Selects the attenuator mode, when the RF signal is switched off.
The setting of the RF OFF mode is not affected by an instrument preset (PRESET
key),
*RST
and the "Save/Recall" function. This parameter is influenced only by the
Parameters:
<Mode>
UNCHanged | FATTenuation
UNCHanged
Freezes the setting of the attenuator when RF is switched off.
The attenuator is only activated when RF is switched on.
This setting recommended if a constant VSWR (
V
oltage
S
tand-
ing
W
ave
R
atio) is required.
FATTenuation
Sets attenuation to maximum when the RF signal is switched off.
This setting is recommended for applications that require a high
level of noise suppression.
*RST:
depends on instrument hardware
Example:
SOUR:POW:ATT:RFOF:MODE FATT
sets the RF OFF attenuator to maximum.
Manual operation:
See
[:SOURce<hw>]:POWer:EMF:STATe
<State>
Displays the signal level as voltage of the EMF. The displayed value represents the
voltage over a 50 Ohm load.
SOURce Subsystem
Summary of Contents for SMJ100A
Page 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Page 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Page 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Page 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Page 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...