Example:
SOUR:LIST:TRIG:EXEC
triggers the processing of the selected list.
Usage:
Event
Manual operation:
See
[:SOURce<hw>]:LIST:TRIGger:SOURce
<Source>
Selects the trigger source processing lists.
The names of the parameters correspond to those under sweep mode. SCPI uses
other names for the parameters; these names are also accepted by the instrument.
The SCPI names should be used if compatibility is an important consideration. An
overview of the various names is given in the following table:
R&S name
SCPI name
Command under manual con-
trol
AUTO
IMMediate
MODE AUTO
SINGle
BUS
MODE SINGLE
or
STEP
EXTernal
EXTernal
MODE EXT TRIG SINGLE
or
EXT TRIG STEP
Parameters:
<Source>
AUTO | IMMediate | SINGle | BUS | EXTernal
AUTO|IMMediate
The trigger is free-running, i.e. the trigger condition is fulfilled
continuously. The selected list in List mode is restarted as soon
as it is finished.
SINGle|BUS
The list is triggered by the GPIB commands
EXTernal
The list is triggered externally via the INST TRIG connector. The
list is executed once.
*RST:
AUTO
Example:
LIST:TRIG:SOUR EXT
selects triggering by means of the external trigger.
Manual operation:
See
7.16.12 SOURce:MODulation Subsystem
This subsystem contains the command for switching on/off all modulations.
[:SOURce<hw>]:MODulation[:ALL][:STATe]
<State>
Activates/deactivates the modulations.
SOURce Subsystem
Summary of Contents for SMJ100A
Page 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Page 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Page 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Page 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Page 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...