Remote Control Commands
R&S
®
SMJ100A
720
Operating Manual 1403.7458.32 ─ 14
Note:
This value also applies to the step width of the rotary knob of the instrument and
increases or decreases the frequency accordingly, when you work in user-defined step
mode.
Parameters:
<Increment>
float
Range:
full frequency range
Increment: see the data sheet: RF characteristics > Resolution
of setting
*RST:
1E6
Example:
FREQ:STEP 50 kHz
sets the step width for the frequency setting to 50 kHz.
Manual operation:
See
[:SOURce<hw>]:FREQuency:STEP:MODE
<Mode>
Activates (
USER
) or deactivates (
DECimal
) the user-defined step width used when
varying the frequency value with the frequency values
UP/DOWN
. The command is
linked to the command "Variation Active" for manual control, i.e. the command also
activates/deactivates the user-defined step width used when varying the frequency
value with the rotary knob.
Parameters:
<Mode>
DECimal | USER
*RST:
DECimal
Example:
FREQ:STEP 50 kHz
sets the step width for the frequency setting to 50 kHz.
FREQ:STEP:MODE USER
actives this step width for frequency variation with the rotary
knob (manual control) and with frequency values
UP/DOWN
(remote control).
Manual operation:
See
7.16.7 SOURce:INPut Subsystem
The
SOURce:INPut
subsystem contains the commands for configuring the inputs for
trigger, data and control signals. A common trigger threshold and input impedance is
effective for all trigger and control signal inputs.
The settings influence the digital modulations, the generation of waveforms or multicar-
rier signals, and all digital standards. Irrespective of this, a common threshold and
input impedance is effective for the serial and parallel data input (Path A only). These
data sources are available for digital modulation ("Custom Digital Modulation").
[:SOURce]:INPut:DATA:IMPedance
.................................................................................721
[:SOURce]:INPut:DATA:THReshold
................................................................................. 721
[:SOURce<hw>]:INPut:MODext:IMPedance
......................................................................721
[:SOURce]:INPut:TRIGger:BBANd:SLOPe
........................................................................722
SOURce Subsystem
Summary of Contents for SMJ100A
Page 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Page 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Page 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Page 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Page 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...