Remote Control Commands
R&S
®
SMJ100A
523
Operating Manual 1403.7458.32 ─ 14
Example:
CAL:IQM:FULL?
starts adjustments for the I/Q modulator across the entire fre-
quency range.
Response: "0"
adjustment has been performed successfully
Usage:
Query only
Manual operation:
See
"Adjust I/Q Modulator Full Range"
:CALibration<hw>:IQModulator:LOCal?
Starts adjustment of the I/Q modulator for the currently set frequency, I/Q swap and
baseband gain. The I/Q modulator is adjusted with respect to carrier leakage, I/Q
imbalance and quadrature.
Return values:
<Local>
0 | 1
*RST:
0
Example:
CAL:IQM:LOC?
starts adjustment for the I/Q modulator for the currently set fre-
quency.
Response:
0
adjustment has been performed successfully
Usage:
Query only
Manual operation:
See
"Adjust I/Q Modulator Current Frequency"
:CALibration<hw>:LEVel[:MEASure]?
[<Force>]
Starts all adjustments which affect the level.
Query parameters:
<Force>
string
*RST:
force
Return values:
<Measure>
0 | 1
Example:
CAL:LEV:MEAS?
starts adjustments for maximum level accuracy.
Response: "0"
adjustment has been performed successfully.
Usage:
Query only
Manual operation:
See
:CALibration<hw>:LEVel:STATe
<State>
Activates/deactivates internal level correction.
CALibration Subsystem
Summary of Contents for SMJ100A
Page 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Page 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Page 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Page 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Page 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...