Remote Control Commands
R&S
®
SMJ100A
507
Operating Manual 1403.7458.32 ─ 14
Figure 7-8: I and Q values calculation and conversion
3. Use an ASCII editor which is able to handle binary data. Create and add the follow-
ing mandatory tags before this binary data set can be further processed:
●
●
●
An example of the
SICO.wv
file contents could be:
{TYPE: SMU-WV, 0}{CLOCK: 10e6}{LEVEL OFFS: 3.010300,0.000000}
0000FF7F8E27BB79 ... 72D8BB79
To simplify the example, the checksum is set to 0, i.e. the instrument does not eval-
uate a checksum.
Tip:
The tags
TYPE
,
CLOCK
,
LEVEL OFFS
and
WAVEFORM
are mandatory for each
waveform. All other tags are optional and can be inserted after the
TYPE
tag in
arbitrary order.
4. Pack the binary data into a
tag with the described structure.
{WAVEFORM-Length: #I0Q0I1Q1I2Q2 ... InQn}
● Calculate the
Length
Length = Number of I/Q pairs * 4 + 1 = 20*4 + 1 = 81 bytes
● Place the string
{WAVEFORM-81:#
at the beginning of the data set
● Place the symbol
}
at the end of the data set
The contents of the waveform file
SICO.wv
for 20 I/Q pairs is now ready for opera-
tion and reads:
{TYPE: SMU-WV,0}
{CLOCK: 10e6}
{LEVEL OFFS: 3.010300,0.000000}
{WAVEFORM-81:#I0Q0I1Q1...InQn}
Waveform and List Format
Summary of Contents for SMJ100A
Page 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Page 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Page 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Page 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Page 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...