![R&S SMJ100A Operating Manual Download Page 501](http://html1.mh-extra.com/html/rands/smj100a/smj100a_operating-manual_3086611501.webp)
Remote Control Commands
R&S
®
SMJ100A
485
Operating Manual 1403.7458.32 ─ 14
Usage:
Query only
*IDN?
Identification
Returns the instrument identification.
Return values:
<ID>
"Rohde&Schwarz,<device type>,<serial number>,<firmware ver-
sion>"
Example:
Rohde&Schwarz,SMJ,
1407.6004k02/000000,3.1.17.1-03.01.158
Usage:
Query only
Manual operation:
See
"Hardware Options / Software Options"
*IST?
Individual status query
Returns the contents of the IST flag in decimal form. The IST flag is the status bit
which is sent during a parallel poll.
Return values:
<ISTflag>
0 | 1
Usage:
Query only
*OPC
Operation complete
Sets bit 0 in the event status register when all preceding commands have been execu-
ted. This bit can be used to initiate a service request. The query form writes a "1" into
the output buffer as soon as all preceding commands have been executed. This is
used for command synchronization.
*OPT?
Option identification query
Queries the options included in the instrument. For a list of all available options and
their description refer to the data sheet.
Return values:
<Options>
The query returns a list of options. The options are returned at
fixed positions in a comma-separated string. A zero is returned
for options that are not installed.
Usage:
Query only
Common Commands
Summary of Contents for SMJ100A
Page 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Page 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Page 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Page 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Page 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...