Instrument Function
R&S
®
SMJ100A
142
Operating Manual 1403.7458.32 ─ 14
Threshold Data Input
Sets the high/low threshold in volts for the serial and parallel data input.
The serial data is input at the DATA connector of the instrument. The parallel data is
input via the DATA pins of the AUX I/O interface.
Remote command:
[:SOURce]:INPut:DATA:THReshold
Threshold Trigger/Control Input
Sets the high/low threshold in volts for the trigger and control signal inputs of the base-
band section.
The setting affects the TRIGGER 1 and TRIGGER 2 inputs (BNC connectors) and the
CW, BURST and LEV_ATT inputs of the AUX I/O interface.
Remote command:
Impedance Data Input
Selects the input impedance for the serial and parallel data inputs.
50 Ohm/GND should be selected for high clock rates.
The serial data is input at the DATA connector of the instrument. The parallel data is
input via the DATA pins of the AUX I/O interface.
Remote command:
[:SOURce]:INPut:DATA:IMPedance
Impedance Trigger/Control Input
Selects the input impedance for the external trigger and control signal inputs.
50 Ohm/GND should be selected for high clock rates.
The setting affects the TRIGGER 1 and TRIGGER 2 inputs (BNC connectors) and the
CW, BURST and LEV_ATT inputs of the AUX I/O interface.
Remote command:
[:SOURce]:INPut:TRIGger:IMPedance
General Instrument Settings
Summary of Contents for SMJ100A
Page 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Page 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Page 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Page 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Page 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...