Trying out the instrument
R&S
®
SMCV100B
48
Getting Started 1432.7046.02 ─ 04
6.4
Enabling and configuring a marker signal
Test setups often require synchronization of an external device with the generated
data stream. For this purpose, the R&S
SMCV100B can output maximum two
marker signals (or markers) also to the generated signal.
The R&S
SMCV100B provides four regular marker signals. You can output two of
the marker signals, one to each of the "User 1"/"User 2" connectors.
With suitable marker settings for instance, you can mark slot or frame boundaries
or mark the start of a particular modulation symbol.
Prerequisites
●
●
Option custom digital modulation R&S
SMCVB-K199
This example extends further the configurations performed in
erating a digitally modulated signal"
on page 40. We assume a default connector
This test setup requires one oscilloscope, like the R&S
®
RTO, as additional equip-
ment.
1. In the block diagram, select "Block Diagram > Baseband > Misc > Custom
Digital Mod > Marker" tab.
2. Select "Marker Mode > Marker 1 > Pulse" and "Divider = 32".
Generated is a periodic marker with marker frequency of 120
KHz. The signal
is output at the User 1 connector of the R&S
SMCV100B (see
3. Use a suitable cable to connect the User 1 connector of the R&S
SMCV100B
to the monitoring instrument, for example an oscilloscope like R&S
®
RTO. See
.
Enabling and configuring a marker signal