Preparing for Use
R&S
®
SMB100B
18
Getting Started 1422.1045.02 ─ 03
To mount the R&S
SMB100B in a rack
1. Use an adapter kit that fits the dimensions of the R&S
SMB100B to prepare
the instrument for rack mounting. For information on the dimensions, see data
sheet.
a) Order the rack adapter kit designed for the R&S
SMB100B. For the order
number, see data sheet.
b) Mount the adapter kit. Follow the assembly instructions provided with the
adapter kit.
2. Lift the R&S
SMB100B to shelf height.
3. Grab the handles and push the R&S
SMB100B onto the shelf until the rack
brackets fit closely to the rack.
4. Tighten all screws at the rack brackets with a tightening torque of 1.2
Nm to
secure the R&S
SMB100B at the rack.
To unmount the R&S
SMB100B from a rack
1. Loosen the screws at the rack brackets.
2. Bring the lifting equipment to shelf height.
3. Remove the R&S
SMB100B from the rack.
4. If placing the R&S
SMB100B on a bench top again, unmount the adapter kit
from the R&S
SMB100B. Follow the instructions provided with the adapter kit.
4.5
Important Aspects for Test Setup
Cable selection and electromagnetic interference (EMI)
Electromagnetic interference (EMI) can affect the measurement results.
To suppress electromagnetic radiation during operation:
●
Use high-quality shielded cables, especially for the following connector types:
– BNC
Double-shielded BNC cables.
– USB
Double-shielded USB cables.
How to:
Chapter 4.8, "Connecting USB Devices"
Important Aspects for Test Setup