Remote Control Commands
R&S
®
SMB100A
327
Operating Manual 1407.0806.32 ─ 21
[:SOURce<hw>]:AM:SENSitivity?
Queries the input sensitivity of the externally applied signal for amplitude modulation.
The sensitivity depends on the set modulation depth, see
[:SOURce<hw>]:AM:DEPTh:EXPonential
The returned value reports the sensitivity in
%/V. It is assigned to the voltage value for
full modulation of the input.
Return values:
<Sensitivity>
float
Range:
0 to 100
Example:
AM:DEPT 50
sets a modulation depth of 50 %.
AM:SENS?
queries the input sensitivity at the external modulation input.
Response: 50
since the voltage value for full modulation is 1V, the resulting
sensitivity is precisely 50 %/V.
Usage:
Query only
Manual operation:
See
[:SOURce<hw>]:AM:SOURce
<Source>
Selects the modulation signal source for amplitude modulation.
on page 328), you can use both, the
internal and an external modulation signal at a time, for example to perform two-tone
AM.
Parameters:
<Source>
INTernal | EXTernal | INT,EXT
INTernal
Uses the internally generated signal for modulation. To configure
the frequency, use the commands of the
EXTernal
Uses an externally applied modulation signal.
INT,EXT
Uses both, the internal and external modulation signals.
*RST:
INT
Example:
AM:SOUR INT
selects the internal modulation source.
Manual operation:
See
SOURce Subsystem