Remote Control Commands
R&S
®
SMB100A
314
Operating Manual 1407.0806.32 ─ 21
:OUTPut<hw>:IMPedance?
Queries the impedance of the RF outputs. It enables you to convert the output level
units between V and W. The impedances cannot be changed.
Return values:
<Impedance>
G1K | G50 | G10K
*RST:
G50
Default unit: Ohm
Example:
OUTP:IMP
queries the impedance of RF output.
Response: 50
the impedance is 50 ohms
Usage:
Query only
:OUTPut<hw>:PROTection:CLEar
Resets the protective circuit after it has been tripped. The state of the output is again
determined by
OUTPut:STATe
.
Example:
OUTP:PROT:CLE
resets the protective circuit for RF output.
Usage:
Event
Manual operation:
See
:OUTPut<hw>:PROTection:TRIPped?
Queries the state of the protective circuit.
Return values:
<Tripped>
0 | 1 | OFF | ON
*RST:
0
Example:
OUTP:PROT:TRIP
Queries the state of the protective circuit for RF output A.
Response: 0
The protective circuit has not tripped.
Response: 1
The protective circuit has tripped.
Usage:
Query only
Manual operation:
See
:OUTPut<hw>[:STATe]
<State>
Activates and deactivates the RF output signal (RF ON / RF OFF).
OUTPut Subsystem