Instrument Function
R&S
®
SMB100A
176
Operating Manual 1407.0806.32 ─ 21
RF Frequency Sweep Settings
► To access the sweep dialog, select "RF > configure > Sweep/List > RF Frequency
Sweep".
In these dialogs you can configure the corresponding sweep signal.
State - Frequency Sweep
Activates RF sweep mode.
Note:
Activating a sweep mode automatically deactivates other sweeps and the list mode.
Remote command:
Mode - RF Frequency Sweep
Selects the RF frequency sweep mode.
If you change the sweep mode during the execution, the signal generator stops the
sweep and starts with the next trigger event at the initial value.
The "Reset Sweep" button sets the sweep to the start value.
"Auto"
Generates a continuously repeating sweep signal immediately after
activating the sweep mode.
The sweep steps are performed atomatically, controlled by the dwell
time, see
"Dwell Time - Frequency Sweep"
Example:
SOUR:SWE:FREQ:MODE AUTO
TRIG:FSW:SOUR AUTO
SOUR:FREQ:MODE SWE
"Single"
Generates a single sweep cycle after a trigger event.
The sweep steps within the cycle are performed automatically, con-
trolled by the dwell time. If one cycle is completed, the instrument
waits for the next trigger event.
To trigger the sweep, use "Execute Single Sweep" button, or the cor-
responding remote control commands, for example
*TRG
.
Example:
SOUR:SWE:FREQ:MODE AUTO
TRIG:FSW:SOUR SING
SOUR:FREQ:MODE SWE
SOUR:SWE:FREQ:EXEC