Troubleshooting and Error Messages
R&S
®
SMA100B
503
User Manual 1178.3834.02 ─ 03
Error
Code
Error
Description
Remedy
183
Adjustment data invalid
Adjustment data is invalid and must be restored.
The adjustment data has to be generated
again by an internal or external adjust-
ment or to be loaded into the instrument.
200
Cannot access hardware
The data transmission to a module was unsuc-
cessful.
The module is not installed, not properly
installed or missing.
201
Hardware revision out of
date
A later version of certain parts of the instrument is
necessary to execute the function selected.
The driver does not support the installed
version of a module.
202
Cannot access the
EEPROM
An error occurs when writing or reading a
EEPROM.
The EEPROM might be defect and has
to be replaced.
203
Invalid EEPROM data
Reading a EEPROM is possible, however the data
are inconsistent.
204
Driver initialization failed
Initialization of a driver fails when booting the
instrument firmware.
The driver is not compatible with the
hardware or software configuration of the
instrument.
241
No current list
There is no list selected. To execute the required
operation, a list has to be selected in the related
dialog. If no list is available, a new list must be cre-
ated.
242
Unknown list type speci-
fied
The list type selected is not valid for the required
operation.
Check the selected list type.
460
Cannot open file
The selected file cannot be opened.
Check the path and file name.
461
Cannot write file
The file cannot be written.
Check if the file is read-only.
462
Cannot read file
The file cannot be read.
Check if the file contents are compatible
with the file type.
463
Filename missing
The required operation cannot be executed
because the file name is not specified.
Enter file name when creating list.
464
Invalid filename extension
The file extension is not valid for the required
operation.
Check the file extension.
465
File contains invalid data
The selected file contains data that is not valid for
the file type.
The file extension determines the data that is valid
for this file type. If the file extension is changed,
the lists are no longer recognized and the data is
therefore invalid.
Check the file extension.
14.4 Querying Error Messages & Info Key
The R&S
SMA100B monitors the functions performed and automatically detects errors
and irregularities. The instrument displays the messages in the info-line and records
them in the background with a detailed description.
Find details to the system messages in
Chapter 14, "Troubleshooting and Error Mes-
Querying Error Messages & Info Key