Remote Control Commands
R&S
®
SMA100B
334
User Manual 1178.3834.02 ─ 03
Return values:
<Catalog>
string
List of all assemblies; the values are separated by commas
The length of the list is variable and depends on the instrument
equipment configuration.
Example:
See
Usage:
Query only
:DIAGnostic<hw>:BGINfo?
[<Board>]
Queries information on the modules available in the instrument, using the variant and
revision state.
Query parameters:
<Board>
string
Module name, as queried with the command
DIAGnostic<hw>:BGINfo:CATalog?
To retrieve a complete list of all modules, omit the parameter.
The length of the list is variable and depends on the instrument
equipment configuration.
Return values:
<BgInfo>
<Module name> <Module stock number incl. variant> <Module
revision> <Module serial number>
List of comma-separated entries, one entry per module.
Each entry for one module consists of four parts that are separa-
ted by space characters.
Example:
See
Usage:
Query only
Manual operation:
See
:DIAGnostic<hw>:POINt:CATalog?
Queries the test points available in the instrument.
For more information, see R&S
SMA100B Service Manual.
Return values:
<Catalog>
string
List of comma-separated values, each representing a test point
Example:
See
Usage:
Query only
DIAGnostic Subsystem