Remote Control Commands
R&S
®
SMA100B
326
User Manual 1178.3834.02 ─ 03
Suffix:
<hw>
.
[1]
Optional suffix
Setting parameters:
<State>
0 | 1 | OFF | ON
*RST:
0
Example:
CALibration:DEBug 1
Usage:
Setting only
:CALibration<hw>:FMOFfset[:MEASure]?
Starts adjustment of the FM/PhiM modulator.
Suffix:
<hw>
.
[1]
Optional suffix
Return values:
<Measure>
0 | 1 | OFF | ON
Example:
CALibration:FMOFfset:MEASure?
// starts adjustment
// 0
// Adjustment successful
Usage:
Query only
Options:
R&S
SMAB-K720
:CALibration<hw>:FREQuency[:MEASure]?
Performs all adjustments which affect the frequency.
Suffix:
<hw>
.
[1]
Optional suffix
Return values:
<Measure>
0 | 1 | OFF | ON
Example:
// Start adjustments for maximum frequency accuracy
CALibration:FREQuency:MEASure?
// 0
// Adjustments are performed successfully
Usage:
Query only
:CALibration<hw>:LEVel:EXTern:DATA
<Data>
Queries what data has been used for the level calibration.
CALibration Subsystem