Varying the RF Signal in List or Sweep Mode
R&S
®
SMA100B
134
User Manual 1178.3834.02 ─ 03
"Live"
Generates the signal directly from the database.
The instrument reads the pairs of values from the list, calculates the
hardware settings and generates the signal immediately.
See
Remote command:
Trigger Slope
Selects the polarity of the active slope of an applied instrument trigger.
Trigger signal is expected at the INST TRIG connector.
"Positive"
Activates the rising edge of the trigger signal.
"Negative"
Activates the falling edge of the trigger signal.
Remote command:
Execute Single
Manually starts list processing in "Single" mode.
Remote command:
[:SOURce<hw>]:LIST:TRIGger:EXECute
Reset
Resets the list to the starting point.
Remote command:
5.6.2 List Mode Data Settings
Access:
1. Select "Sweep" > "List mode".
2. Select "List Mode Data".
This dialog contains the parameters required for creating and editing lists, activat-
ing the learning function and selecting the list processing mode.
List Mode Settings