Varying the RF Signal in List or Sweep Mode
R&S
®
SMA100B
112
User Manual 1178.3834.02 ─ 03
Table 5-1: Cross-reference between manual and remote control in Auto mode (Sweep/List)
Manual control mode:
"Auto"
Remote commands
Rohde
&
Schwarz proprietary and [SCPI conform] value name
RF frequency sweep
:TRIGger<hw>:FSWeep:SOURce AUTO [IMMediate]
:SOURce<hw>:SWEep:FREQuency:MODE AUTO
RF level sweep
:TRIGger<hw>:PSWeep:SOURce AUTO [IMMediate]
:SOURce<hw>:SWEep:POWer:MODE AUTO
LF frequency sweep
:TRIGger<hw>:LFFSweep:SOURce AUTO [IMMediate]
:SOURce<hw>:LFOutput<ch>:SWEep:FREQuency:MODE AUTO
List
:SOURce<hw>:LIST:MODE AUTO
:SOURce<hw>:LIST:TRIGger:SOURce AUTO
Single / Extern Single mode (Sweep/List)
Figure 5-4: Single / Extern Single mode (sweep / list )
In single mode, you can specify, that the signal returns to the start value when a sweep
cycle has been completed.
See
●
The instrument generates a single sweep cycle.
●
Trigger mode "Manual". A trigger event initiates one sweep from the start value to
the end value.
●
"State = On" sets the signal to the start value: the sweep start frequency, the
sweep start power or the frequency-power value pair of the selected index in the
list.
●
Starts signal generation with a trigger event.
Signal Generation and Triggering in the Sweep and List Modes