Remote Control Commands
R&S
®
SGT100A
484
User Manual 1176.8674.02 ─ 07
Usage:
Query only
Manual operation:
See
:SYSTem:COMMunicate:SOCKet:RESource?
Queries the visa resource string for remote control via LAN interface, using TCP/IP
socket protocol.
Return values:
<Resource>
string
Example:
SYSTem:COMMunicate:SOCKet:RESource?
Response:
"TCPIP::10.113.1.150::5025::SOCKET"
Usage:
Query only
Manual operation:
See
:SYSTem:COMMunicate:USB:RESource?
Queries the visa resource string for remote control via the USB interface.
Return values:
<Resource>
string
Example:
SYSTem:COMMunicate:USB:RESource?
queries the VISA resource string for remote control via the USB
interface.
Response:
"USB::72::000000::INSTR"
Usage:
Query only
Manual operation:
See
:SYSTem:IDENtification
<Identification>
Selects the mode the instrument identification is performed.
Parameters:
<Identification>
AUTO | USER
AUTO
The "IDN String" and the "OPT String" are set automatically.
USER
Enables the selection of user definable "IDN String" and "OPT
String".
*RST:
AUTO
Example:
SYST:IDEN USER
selects the user defined identification string.
Manual operation:
See
SYSTem Subsystem